Elastography of the thyroid nodule, cut-off points between benign and malignant lesions for strain, 2D shear wave real time and point shear wave: a correlation with pathology, ACR TIRADS and Alpha Score

Mena, Glenn and Montalvo, Alejandro and Ubidia, Michael and Olmedo, Julio and Guerrero, Ana and Leon-Rojas, Jose E. (2023) Elastography of the thyroid nodule, cut-off points between benign and malignant lesions for strain, 2D shear wave real time and point shear wave: a correlation with pathology, ACR TIRADS and Alpha Score. Frontiers in Endocrinology, 14. ISSN 1664-2392

[thumbnail of pubmed-zip/versions/1/package-entries/fendo-14-1182557.pdf] Text
pubmed-zip/versions/1/package-entries/fendo-14-1182557.pdf - Published Version

Download (3MB)

Abstract

Objective: A prospective cross-sectional investigation of 170 thyroid nodules (TN) between January 2020 and December 2021 at Alpha Imagen was conducted to determine cut-off points (C/O) for elastography measurements and their diagnostic accuracy.

Methods: Nodules were categorized by ACR TI-RADS, Alpha Score (AS), and Bethesda; all were evaluated using 2D Shear Wave Real Time Elastography (RT-SWE), point Shear Wave (pSWE), and Strain Elastography (SE). Data was assessed with ROC curves, the Shapiro-Wilk test, T test, Chi-square test, and ANOVA.

Results: C/O were as follows: RTSWE Emax of 115kPa and 6.5 m/s, Emean of 47.5 kPa and 4.1 m/s, pSWE (average) of 52.4 kpa and 4.15 m/s; sensitivity of 81.2% and specificity of 57.6%, with a PPV of 72.4% and NPV of 70.0%. SE Value A had a C/O of 0.20%, with a sensitivity of 84%, specificity of 57%, PPV of 72.4% and NPP of 73.6%. The Strain Ratio nodule/tissue C/O was calculated as 2.69, with a sensitivity of 84%, specificity of 57%, PPV of 72.3%, and NPV of 73.5%. The RLBIndex quality control must be at least 92%; for pSWE, we suggest a mean interquartile ratio of ≤15.7% for kPa and 8.1% for m/s. The recommended depth is between 1.2 and 1.5 cm, and commonly used ROI boxes were 3x3 and 5x5mm.

Conclusion: 2D-SWE and pSWE with Emax and Emean demonstrated C/O with excellent diagnostic accuracy. To maximize the correct classification of TN, we suggest combining ACR TI-RADS and AS with any of the elastography measurements assessed here.